N5230C Option 140/145/146/240/245/246
Corrected system performance with 3.5 mm connectors1 (continued)
Table 4. N4433A Electronic calibration module
N5230C
• Option 140 or 240 standard test set and standard power range
• Option 145 or 245 configurable test set and extended power range
• Option 146 or 246 configurable test set and extended power range with internal
second source
Applies to the N5230C Option 140/145/146/240/245/246 analyzers, N4433A
electronic calibration module, 85131F flexible test port cable set, and a full 4-port
calibration. Also applies to the following condition: Environmental temperature
23° ±3 °C, with < 1 °C deviation from calibration temperature.
Specification (dB)
Description
10 to
500 MHz
500 MHz to
2 GHz
2 to
8 GHz
8 to
13.5 GHz
Directivity
52
52
47
45
Source match
42
42
39
31
Load match
45
45
41
35
Reflection
tracking
±0.060
(+0.01/°C)
±0.060
(+0.01/°C)
±0.090
(+0.02/°C)
±0.040
(+0.03/°C)
Transmission
±0.045
±0.039
±0.055
±0.127
tracking
(+0.01/°C)
(+0.01/°C) (+0.02/°C) (+0.03/°C)
Transmission uncertainty (specifications)
13.5 to
20 GHz
45
31
35
±0.180
(+0.03/°C)
±0.160
(+0.03/°C)
Magnitude
N5230C Option 140/145/146/240/245/246 full 2-port2 cal using N4433A
10
■ 10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
1
● 8 to 20 GHz
0.1
0.01
10
S11 = S22 = 0
Source power = -5 dBm
0 -10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient (dB)
Phase N5230C Option 140/145/146/240/245/246 full 2-port2 cal using N4433A
100
■ 10 to 500 MHz
♦ 500 MHz to 2 GHz
▲ 2 to 8 GHz
10 ● 8 to 20 GHz
1
0.1
10
S11 = S22 = 0
Source power = -5 dBm
0 -10 -20 -30 -40 -50 -60 -70 -80 -90
Transmission coefficient (dB)
1. From 300 kHz to 10 MHz, performance is characterized as “typical”. To generate these typical values,
please download our free Uncertainty Calculator from www.agilent.com/find/na_calculator.
8
2. All of the curves are for 2-port calibrations. Multiport uncertainties are currently not supported.