DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

C9329 데이터 시트보기 (PDF) - Hamamatsu Photonics

부품명
상세내역
제조사
C9329
Hamamatsu
Hamamatsu Photonics 
C9329 Datasheet PDF : 3 Pages
1 2 3
Photosensor amplifier C9329
s Typical connection to photodiode
This is an example using a photodiode whose cathode is
internally connected to its metal package.
When you use a photodiode metal package, use an insu-
lator to electrically insulate and also hold the package in
a shield case as shown in the figure at right. Connect the
anode to the shield case.
Any single-element photodiode with a terminal capaci-
tance below 5000 pF can be used.
Using a photodiode with anode grounded is recom-
mended.
Using a photodiode with a BNC connector (S2281 se-
ries) allows you to easily make measurements because it
connects to C9329 with a BNC-BNC plug coaxial cable.
PHOTODIODE
ANODE
SOLDER
Connect to C9329 terminal
CATHODE
COAXIAL CABLE WITH BNC PLUG
INSULATOR EXTERNAL SHIELD CASE
KACCC0221EA
Anode: Connect to the shield wire of the cable and shield case.
Cathode: Connect to the core wire of the cable.
s Connection example
Operation example by digital output (REMOTE MODE)
s Display example of accessory sample software
COAXIAL CABLE
WITH BNC-BNC PLUG RS-232C CABLE
PC
PHOTODIODE
(S2281 SERIES)
to IN
TERMINAL
to RS-232C
TERMINAL
to DC TERMINAL
SWITCH C9329
REAR SW TO
"REMOTE MODE"
AC ADAPTER
(Need not be connected
when operating on battery)
Rf RANGE SWITCH
KACCC0222EA
Note: Use the Rf RANGE switch to change the detection sensitivity.
(Detection sensitivity cannot be changed from the PC.)
s Dimensional outline (unit: mm, excluding project parts)
90 ± 1
FRONT
Rf RANGE ZERO ADJ. DATA LOG.
H
AUTO ZERO
POWER
ON
M
L
PUSH START
PHOTOSENSOR AMPLIFIER C9329
REAR
MANUAL
REMOTE
MODE
MODE
OUT
IN
UPPER SURFACE
Data logger setting range
Measurement interval: 50 ms to 1 min
(50 ms interval)
Measurement count: 32000 Max.
Measurement interval × Meassurement count:
20 hours Max.
DC JACK
RS232C
DC12V EIAJ ONLY FOR REMOTE MODES
s Accessories
AC adapter (Plug type: A-2 plug) *5
Dry battery
Sample software CD-ROM (OS: Windows98SE/Me/2000/Xp *6
Instruction manual
KACCA0137EA
*5: Caution) Depending on the country, an adapter plug might be required when connecting to the AC outlet.
If so, please purchase a proper adapter plug from an electronics supply house.
*6: Resistered trademark of Microsoft Corporation in the United States.
Photodiode, coaxial cable with BNC-BNC plug and RS-232C cable are not supplied with C9329. You will need an RS-232C
cable (straight cable terminated with a D-sub 9 pin female connector at both ends) available from electronics supply houses.