DatasheetQ Logo
Electronic component search and free download site. Transistors,MosFET ,Diode,Integrated circuits

HAL1000UT-E 데이터 시트보기 (PDF) - Micronas

부품명
상세내역
제조사
HAL1000UT-E
Micronas
Micronas 
HAL1000UT-E Datasheet PDF : 24 Pages
1 2 3 4 5 6 7 8 9 10 Next Last
HAL 1000
ADVANCE INFORMATION
Step 2: Calculation of the Sensor Parameters
Fig. 29 shows the typical characteristics for a contact-
less switch. There is a mechanical range where the
sensor must be switched high and where the sensor
must be switched low.
VOUT
Sensor
switched off
Hysteresis
(here 80 %)
Sensor
switched on
Calibration points
position
Fig. 29: Characteristics of a position switch
Set the system to the calibration point where the sen-
sor output must be high, and press the key Readout
BOFF. The result is the corresponding ADC-READ-
OUT value.
Note: The magnetic south pole on the branded side
generates negative ADC-READOUT values, the north
polarity positive values.
Then, set the system to the calibration point where the
sensor output must be low, press the key Readout
BONand get the second ADC-READOUT value.
Now, adjust the hysteresis to the desired value. The
hysteresis is the difference between the switching
points and suppresses oscillation of the output signal.
With 100% hysteresis, the sensor will switch low and
high exactly at the calibration points. A lower value will
adjust the switching points closer together. Fig. 29
shows an example with 80% hysteresis.
By pressing the key calibrate and store, the software
will calculate the corresponding parameters for Sensi-
tivity, VOQ, Low-Level, High-Level and store these val-
ues in the EEPROM.
This calibration must be done individually for each sen-
sor.
The sensor is now calibrated for the customer applica-
tion. However, the programming can be changed again
and again if necessary.
Step 3: Locking the Sensor
The last step is activating the LOCK function with the
lockkey. The sensor is now locked and does not
respond to any programming or reading commands.
Warning: The LOCKR register cannot be reset!
2.5. Example: Calibration of a Position Switch
The following description explains the calibration pro-
cedure using a position switch as an example.
The mechanical switching points are given.
temperature coefficient of the magnet: 500 ppm/K
Step 1: Input of the registers which need not be
adjusted individually
The register values for the following registers are given
for all sensors in the application:
FILTER
Select the filter frequency: 500 Hz
RANGE
Select the magnetic field range: 30 mT
TC
For this magnetic material: 6
TCSQ
For this magnetic material: 14
Enter these values in the software, and use the write
and storecommand to store these values perma-
nently in the registers.
Step 2: Calculation of the Sensor Parameters
Set the system to the calibration point where the sen-
sor output must be high, and press Readout BOFF.
Set the system to the calibration point where the sen-
sor output must be low, and press Readout BON
Now, adjust the hysteresis to 80%, and press the key
calibrate and store.
Step 3: Locking the Sensor
The last step is activating the LOCK function with the
LOCKcommand. The sensor is now locked and does
not respond to any programming or reading com-
mands.
Warning: The LOCKR register cannot be reset!
10
Micronas

Share Link: 

datasheetq.com  [ Privacy Policy ]Request Datasheet ] [ Contact Us ]