Corrected System Performance
The specifications in this section apply for measurements made with the N5230C
Options 140, 145, 146, 240, 245, and 246 analyzers with the following conditions:
• 10 Hz IF bandwidth
• No averaging applied to data
• Isolation calibration with an averaging factor of 8
Table 1. System dynamic range at test port1
Standard configuration and standard power range (Options 140, 240)
Description
Specification (dB)
at test port
Typical (dB)
at test port
300 kHz to 10 MHz2
111
10 MHz to 4 GHz2
120
128
4 to 6 GHz
118
129
6 to 10.5 GHz
115
127
10.5 to 13.5 GHz
107
119
13.5 to 15 GHz
107
119
15 to 20 GHz
103
116
Configurable test set and extended power range (Options 145, 245)
Configurable test set, extended power range, and internal second source
(Options 146, 246)
Description
Specification (dB)
at test port
Typical (dB)
at test port
300 kHz to 10 MHz2
111
10 MHz to 4 GHz2
120
128
4 to 6 GHz
118
128
6 to 10.5 GHz
113
125
10.5 to 13.5 GHz
105
117
13.5 to 15 GHz
105
117
15 to 20 GHz
98
115
1. The system dynamic range is calculated as the difference between the noise floor and the specified
source maximum output power. The effective dynamic range must take measurement uncertainties
and interfering signals into account.
2. May be degraded typically by 10 dB at particular frequencies (multiples of 5 MHz) below 500 MHz due
to spurious receiver residuals. Methods are available to regain the full dynamic range.